VNA
The VNA program
has most of the functionality of a modern VNA, program functionality
exceeds regular instruments in:

Operation:
Once the operator has selected start and stop frequency, number of
frequency points, parameter and display settings (either manually or
through the File/Open function) a sweep can be made by pressing the
trigger button. The Continuous/Manual button in the Continuous
position will iteratively measure S-parameters when the trigger is
also pressed, the operator may work on the device to optimize the
reponses. No menu functions will work in continuous mode.
Instrument
Configuration:
Configuration is by 2 means, the configuration file
“C:\e_factory_LV\e_factory_LV.conf”
defines the type of VNA being addressed . These fields may be
programed (with simple text editor) as in the following entry:
- [vna]
-
type=2 # 0=NULL/dummy, 1=HP8510, 2=8720, 3=HP 8410
GPIB configuration is done through the National Instruments
NI488.2 tools. Using “ibconf”
define an instrument “VNA” for the appropriate GPIB
address.
Database
Configuration:
Main
Interface Menus:
File
- Open: Load parameter programming. A window will
pop up with the database-stored devices, there is no “save”
functionality from the VNA
program, device parameters must be programmed through
“db_programmer”.
-
Remote Mode: Program will respond to remote commands through
TCP/IP. This is the means for running custom programs.
-
Save Response: S parameter response will be saved as a
tab-delimited user-named ASCII file. Results may be viewed with a
text editor or imported into a spreadsheet.
-
Post to DB: S parameter response will be saved into
relational DB, operator will be prompted for device identifying
information.
-
Post Cal to DB: Once enabled with SOLT calibration, data will
be posted to DB for automatic loading on “Open”
-
Recall Cal: Load existing calibration from DB
-
Exit: Terminate program.
Parameter
- S11, S12, S22, S21:
S-parameter.
Calibrate
- Off: Raw, uncorrected data.
-
Response: Simpled response division, most useful
for S12 and S21.
-
Instrument: Use native instrument calibration,
places instrument in LOCAL
mode so that operator may calibrate or select an existing instrument
calibration. Once instrument is calibrated, VNA program will retake
remote control of instrument.
-
Programmed: Use DB-programmed calibration and
standards. If properly programmed calibration exists, a sub-menu
will appear for each cal standard.
Mode
- Step: Synthesizer step mode.
-
Ramp: Generator ramp mode.
Plot
- Magnitude: Magnitude in DB of S-parameter.
-
Phase: Phase in degrees of S-parameter.
-
Smith Chart: Traditional Smith Chart representation
of S-parameter.
-
Time Domain: DB Magnitude response in time domain. TDR
parameters defined in TDR section of user panel.
-
Resonances: Front panel table of response Qs.
Marker
- Add: Add frequency marker, note doesn't affect
programmed markers or resonance markers.
-
Del: Delete frequency marker.
-